9
7
4. This procedure completes the setting to the
Multi-Connection mode.
BALANCE (No) Command (PC) RETURN DATA (PC)
a [a] PRINT [a] PRINT 0.0g
b [b] PRINT [b] PRINT 0.0g
c [c] PRINT [c] PRINT 0.0g
d
(nearest to PC)
[d] PRINT 0.0g
(No Data)
Command Codes in the Multi-Connection Mode
Only the commands shown below are valid in the Multi-Connection mode.
(α is the name of the balances.)
[α]TARE Same as the operation of pressing the [O/T] key of the balance “α”.
[α]T Same as the operation of pressing the [O/T] key of the balance “α”.
[α]POWER Same as the operation of pressing the [POWER] key of the balance “α”.
[α]Q Same as the operation of pressing the [POWER] key of the balance “α”.
[α] PRINT Same as the operation of pressing the [PRINT] key of the balance “α”.
[α]D05 Same as the operation of pressing the [PRINT] key of the balance “α”.
[α]D07 Same as sending the D07 command to the balance “α”.
[α]UNIT Same as the operation of pressing the [UNIT] key of the balance “α”.
[α]CAL Same operation as sending the CAL command to the balance “α”.
[α]UNIT+ Same as the state of holding down the [UNIT] key of the balance “α”.
Format of Output Data from the Balance
The output data from the balance “α” is performed with the following format.
“[α]” data <delimiter>
Format of the load data is the same as the normal output form (Format EB type).
Operation is not guaranteed for the data containing characters, such as printing of date, time,
and calibration document etc.