7. Calibration Mode
(1) How to enter
Turn on the Power while pressing the key, and then press key,
and press Enter key as soon as you selecting the number of platform.
If you don't set the number of platform, it will automatically be moved to Normal
Mode. F13 should be set before entering to CAL mode.
(2) Available keys
Key : Change the set value. Used to increase the set value + 1.
Key : Change the digit of the set value. Used to increase the set value ×10.
Key : Move to the next menu.
(3) Calibration Menu(CAL 1 - CAL 5)
CAL 1 : Maximum Capacity Setting
CAL 2 : Minimum Division Setting
CAL 3 : Setting Weight
CAL 4 : Zero Calibration
CAL 5 : Span Calibration