8. Error message and Trouble shooting
Err 02
Reason ▣
Load cell connection failure or error in A/D conversion part.
Trouble shooting☞
Check the load cell connector to see if the polarity of signal is
reversed.
Err 06
Reason ▣
Error in printer connection
Trouble shooting☞
Check with printer connector
If there is no problem with printer connector, please request
A/S to head office.
Err 13
Reason▣
The zero range deviates from the set range.
Trouble shooting☞
Confirm that there is nothing on the weighing platform.
If there is nothing exist, do calibration in CAL mode.
Over
Reason ▣
The weight on platform is too heavy to be measured.
Trouble shooting☞
Do not load cell item exceeds the maximum tolerance.
If the load cell is damaged, the load cell should be replaced.
Err 21
Reason▣
The resolution is set to be exceeded the limit 1/10,000.
Trouble shooting☞