8. Error message and Trouble shooting
Err 02
Reason
Load cell connection failure or error in A/D conversion part.
Trouble shooting
Check the load cell connector to see if the polarity of signal is
reversed.
Err 06
Reason
Error in printer connection
Trouble shooting
Check with printer connector
If there is no problem with printer connector, please request
A/S to head office.
Err 13
Reason
The zero range deviates from the set range.
Trouble shooting
Confirm that there is nothing on the weighing platform.
If there is nothing exist, do calibration in CAL mode.
Over
Reason
The weight on platform is too heavy to be measured.
Trouble shooting
Do not load cell item exceeds the maximum tolerance.
If the load cell is damaged, the load cell should be replaced.
Err 21
Reason
The resolution is set to be exceeded the limit 1/10,000.
Trouble shooting