Unit Normalization
1. Main Menu
2. Press 7 – SMART Trac
3. Press 7 – Unit Normalization
Note: Methods can not be created or imported until unit
normalization is performed.
4. Select SMART Trac II. Press 1 to toggle between SMART Trac I & II.
5. Press Ready
6. Remove sample or oil standard if present on the magnet.
7. Press Ready.
Note: Process will take about 5mins.
8. Collecting data at gain 10, 25 and 50 will be shown on the screen.