EDGE™
Energized Dispersive Guided Extraction
Manual

Table of Contents

Questions and Answers

  • V
    Victoria GarciaAug 1, 2025
    What to do if the Solvent/Extract Recovery is low in CEM EDGE system?
    • D
      David BrownAug 1, 2025
      Low solvent or extract recovery in your CEM Laboratory Equipment may be due to empty solvent bottles. Ensure that the solvent bottles are adequately filled by adding solvent. It could also be caused by slow sample draining. To resolve this, increase the pore size of the Q-Disc. If you are working with cellulose, increase the C#.
  • K
    Karen GuzmanAug 5, 2025
    How to fix a Possible System Clog in CEM EDGE system Laboratory Equipment?
    • D
      Daniel LeeAug 5, 2025
      A possible system clog in your CEM Laboratory Equipment can manifest as slow sample draining. Try increasing the pore size of the Q-Disc. For cellulose, increase C#. Another cause could be overfilled Q-Cups. If this is the case, clean the actuator. Check the “Routine Maintenance” and “Sample Preparation” sections for instructions and preventative measures.
  • R
    Richard JenkinsAug 9, 2025
    What to do if Max Pressure is Exceeded in CEM EDGE system?
    • L
      Lisa McdonaldAug 9, 2025
      If the maximum pressure is exceeded with your CEM Laboratory Equipment, it's likely that the pressure generated at the current temperature exceeds the system's maximum pressure. To resolve this, decrease the method temperature or decrease the amount of sample being used.
  • D
    david85Aug 14, 2025
    Why are recoveries low with my CEM EDGE system Laboratory Equipment?
    • D
      dawn02Aug 14, 2025
      If you are experiencing low recoveries with your CEM Laboratory Equipment, it could be due to a hold time that is not long enough. Try increasing the hold time by 1 minute. Alternatively, the temperature may be incorrect. In this case, increase the temperature by 10 °C, or decrease it by 10 °C if you are working with temperature-sensitive analytes.
  • W
    William AndradeAug 18, 2025
    Why is the Q-Cup sticking to the actuator in my CEM Laboratory Equipment?
    • J
      Joshua JohnsonAug 18, 2025
      If the Q-Cup is sticking to the actuator in your CEM Laboratory Equipment, it might be because the Q-Cups are dirty. Try cleaning the Q-Cups. Another possible cause is that the Q-Cups are overfilled. In this case, clean the actuator. See “Routine Maintenance” for instructions. See “Sample Preparation” for preventative measures.
  • A
    April MooreAug 22, 2025
    How to remove Particulates in extract from CEM EDGE system?
    • D
      Dale BallardAug 22, 2025
      If you find particulates in the extract from your CEM Laboratory Equipment, it may be due to the porosity of the samples being less than that of the Q-Disc. Try decreasing the pore size of the Q-disc. For cellulose, decrease C#.
  • T
    Tonya PetersenAug 27, 2025
    What to do if there is Water in the extract in CEM EDGE system Laboratory Equipment?
    • M
      melissanguyenAug 27, 2025
      If there is water in the extract from your CEM Laboratory Equipment, the sample is likely wet. To resolve this, increase the amount of drying agent used.

Summary

Safety

Safety Notations

Explains the meaning and use of safety alert words like NOTE, CAUTION, and WARNING.

Important Safeguards

Instrument Safeguards

Outlines crucial safety measures for operating and maintaining the EDGE instrument.

Sample Preparation Safeguards

Details safety precautions specifically for sample preparation procedures.

Heating Safeguards

Lists safety guidelines related to the system's heating capabilities and temperature settings.

Tools

Users

System Installation

Unboxing

Instructions for inspecting the EDGE system and its accessories upon delivery.

System Setup: Initial Connections

Step-by-step guide for connecting the exhaust, power, and installing the barcode scanner.

System Setup: Bottle & Waste Lines

Steps for attaching solvent bottles and waste lines to the EDGE system.

System Setup: System Recovery

Instructions for system recovery and autosampler homing procedures.

System Setup: Settings & Rack

Steps for syringe emptying, rack insertion, and initial system settings customization.

System Setup: Calibration Verification

Guide to setting language, date/time, and verifying autosampler calibration.

System Setup: Waste & Autosampler Calibration

Details on verifying autosampler and waste calibration procedures.

System Setup: Pressure Calibration

Steps for performing pressure calibration and completing system setup.

System Setup: Finalization

Final steps for pressure calibration and confirmation of system setup completion.

Sample Analysis

Sample Preparation

Detailed instructions for preparing samples using Q-Cups and Q-Discs.

Prepare the EDGE

Steps for loading sample racks and verifying bottle positions for analysis.

Load Method

Guide for selecting and assigning analytical methods to sample positions.

Run Sample

Instructions for starting sample analysis runs and monitoring their progress.

Run Sample: Details

Details on assigning sample IDs, starting runs, and viewing run status.

Bottle Setup & Configuration

Q-Discs

Enter Q-Disc Code

Steps to input Q-Disc barcodes for sample analysis.

Methods

Create Method

Comprehensive guide for creating custom analytical methods for the EDGE.

Edit Method

Instructions for modifying existing analytical methods.

Data Management

Export Data

Instructions for exporting sample and run data to a flash drive.

Calibrations

Autosampler Calibration: Overview

Detailed instructions for performing a full calibration of the autosampler system.

Autosampler Calibration: Tool Usage

Continues the detailed steps for autosampler calibration, including tool usage.

Autosampler Calibration: Alignment

Further instructions for autosampler calibration, focusing on alignment and drop-off points.

Utilities

Software Update

System Repacking

Routine Maintenance

Troubleshooting

System Specifications

CEM EDGE system Specifications

General IconGeneral
BrandCEM
ModelEDGE system
CategoryLaboratory Equipment
LanguageEnglish