EasyManua.ls Logo

Chattanooga INTELECT LEGEND Series - Premodulated

Chattanooga INTELECT LEGEND Series
63 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21
Intelect® Legend Series Combo
OPERATION
Premodulated
Premodulated is an amplitude modulated sine wave. This waveform is similar
to the beat
frequency created by Interferential current. In some cases, Premodulated therapy provides a
good alternative for Interferential treatment especially when treating areas of the body where
four electrodes can not be utilized.
Cycle Time parameter controls the on/off cycle time of the current. There are 2 available
options, Continuous (default) and 5/5.
When you highlight Beat Frequency and press the enter button, 5 options are displayed.
They are 1-10 Hz, 80-150 Hz (default), 1-150 Hz, Variable and Fixed.
The Variable option allows you to select a Low Beat frequency from 1-200 Hz and a High Beat
frequency from 1-200 Hz. To make changes in the Variable frequency, highlight variable and
press the enter button. To change the Beat Low frequency, use the down arrow to highlight
Beat Low and press enter. Use the up/down arrows to adjust the frequency. Press enter to
accept.
Beat Low describes the lowest frequency in the range of a sweep mode. For example when
using a sweep of 80-150 Hz, 80 Hz is the lowest frequency.
Beat High describes the highest frequency in the range of a sweep mode. For
example when
using a sweep of 80-150 Hz, 150 Hz is the highest frequency.
The Fixed Option, allows you to select a fixed frequency from 1-200 Hz. To make changes to
the Fixed frequency, highlight fixed and press enter. Use the up/down arrows to adjust
frequency. Press enter to accept.