TROUBLESHOOTING
Intelect® RPW
73
ERRORS CONTINUED
Error
Code
Description
118 A button was pressed that accessed the MMC card (e.g., Radial
Pressure Wave Rationale, Contraindications, etc.), but the MMC
inserted into the MMC port is not the correct MMC.
119 The Delete button was pressed while a default (factory) protocol
was highlighted.
121 The START button was pressed, but a handpiece cable is not
connected to any output socket.
122 The START button was pressed, but a handpiece cable is not
connected to the Channel 1 output socket.
123 The START button was pressed, but a handpiece cable is not
connected to the Channel 2 output socket.
124 The user is attempting to use the wrong transmitter for the
selected protocol.
125 The transmitter in Channel 1 has reached one million pulses.
126 The transmitter in Channel 2 has reached one million pulses.
127 The handpiece cable has become disconnected from the
output socket during a therapy session.
In the event that an Error message or Warning appears beginning with
a 2 or 3, immediately stop all use of the unit and contact the dealer or
DJO for service. Errors and Warnings in these categories indicate an
internal problem with the unit that must be tested by DJO or a Field
Service Technician certified by DJO before any further operation or use
of the system.
Use of a unit that indicates an Error or Warning in these categories
may pose a risk of injury to the patient, user, or extensive internal
damage to the system.
In case of display failure or other obvious defects, switch the unit
off immediately by means of the Power On/Off button and notify a
certified service technician.