8.1 Care
Warning!
Shock hazard - Remove the power cord of the Tablet from the wall outlet before cleaning.
Shock hazard, equipment damage -
- Liquids must not enter the device and it’s components, incl. the Tablet. If liquids have
entered into the components, the Revolution Wireless must be immediately checked by
a service technician, before it can be reused.
- Never dismantle the remote control, the modules, the docking station or the AC adapter
as they contain high-voltage parts with a risk of electric shock.
Warning!
Patient hazard - patient contamination
- Before using the unit on another patient, clean and disinfect it according to the
instructions in this section.
Caution!
Equipment damage -
- The plastic material used is not resistant to mineral acids, formic acid, phenols, cresols,
oxidants and strong organic or inorganic acids with a pH value below 4.
- Use only clear disinfectants to prevent discoloration of the device.
- Do not expose the therapy unit to strong ultraviolet radiation (sunlight) and fire.
- Do not sterilize the stimulator.
- Do not immerse in liquids.
Remote Control, Modules, and Docking Station:
All parts of the Revolution Wireless can be disinfected by wiping down with a disinfectant. Thus, it
complies with the special hygiene standards for medical technical equipment.
All components can be cleaned with common disinfectants and mild household detergents.
Only use a soft cloth and an alcohol-based, solvent-free cleaning product, to wipe the therapy unit
down.
Allow the device to completely dry before use.
Electrodes:
Clean electrode surface after each use by holding under dripping water for 5 to 10 seconds while gently
rubbing gel surface with fingertips.
After cleaning, return electrodes to the liner and store in package at room temperature.
Electrode longevity varies, depending on maintenance, skin condition, frequency of usage, type and site
of stimulation. Eective adhesion may be extended by careful cleaning and storing.