7.0DESCRIPTION OF MEASURING MODES
The TI-25M can be used in either the Single Thickness Readingmode or the Scan
mode.
The Single Thickness Readingmode is suitable for most applications. While the
probe remains in contact with the material being measured, the TI-25M performs
four (4) measurements every second, updating the LCD display after each read-
ing. When the probe is removed from the surface, the last reading is retained on
the display.
The Scanmode is used to survey or examine larger sections of material. When
set for Scanmode, the TI-25M performs sixteen (16) measurements every
second, but does not display each reading. This mode is designed for those
applications where the probe will be “dragged” or lightly “scrubbed” over the
measuring surface. While the probe maintains contact with the material being
measured, the TI-25M is keeping track of the lowest (thinnest) measurement it
encounters. Any brief interruptions in the signal will be ignored. When the probe
is removed from the surface for more than one second, the TI-25M will display
the thinnest reading encountered during the measurement survey.
To enable and disable the Scan mode:
Press the key.
Each time the key is pressed the Scan mode status will change from
Off-to-On-to-Off-etc. As the mode is changed, the current status “Off” or “On”
will momentarily flash on the display.