11-3 Electrode Components
Membrane
Dissolved oxygen electrode(Cable length Approx. 5m)
Internal uid inlet
Membrane(Replacement)
Internal uid 30mL(0.5M NaOH)
Dropping pipet 1mL(Internal uid for injection)
11-4 How to Inject Solution
Separate the membrane from the electrode.
Fill the membrane with 1mL of 0.5M NaOH using a dropping pipette.
Assemble the membrane onto the electrode.
•
Be sure not to tighten the membranermly but let it be a little loose so that the
solution may leak.
Loosen the solution opening using a at-head driver.
Inject the solution through the opening.
Do the injection until the solution oozes between the sensor and the membrane.