Recommended materials for cleaning and disinfecting transducer
Solution Origin
Purpose
WIP'ANIOS FRA
Wipe
alcohol/Didecyl
dimethyl ammonium
Disinfection
(
Using clean
cloth soak the liquid
)
Any
Wipe
Clean or
Disinfection
(Using clean cloth soak
)
Any
Wipe
Clean or
Disinfection
Clinell Sporicidal Wipes GBR
Wipe
Percarbonate Citric
Disinfection
mikrozid PAA wipes DEU
Wipe
acid,Hydrogen
Disinfection
Universal wipes GBR
Wipe
Disinfection
perform classic wipes EP DEU
Wipe
Disinfection
CAUTION
These transducers are not designed to withstand heat sterilization methods. Exposure to
temperatures in excess of 60 º C will cause permanent damage. The transducers are not
designed to be totally submerged in fluid, as permanent damage will result if the entire
transducer is submerged.
System Safety
Handling precautions
Ultrasound systems are highly sensitive medical instruments that can easily be damaged by improper
handling. Use care when handling and protect from damage when not in use. DO NOT use a damaged or
defective system. Failure to follow these precautions can result in serious injury and equipment damage.
Electrical shock hazard:
The system is driven with electrical energy that can injure the patient or user if live internal parts are
contacted by conductive solution:
DO NOT immerse the system into any liquid beyond the level indicated by the immersion level diagram.
Never immerse the system connector into any liquid.
Prior to each use, visually inspect the system lens for cracks, cuts, tears, and other signs of physical
damage. DO NOT use a system that appears to be damaged until you verify functional and safe