SDmatic
Verification of the electronic chain
Press once on the small arrow.
The first value is a calibration numerical value. The second value
is an internal resistance measurement (high value). The value of
this resistance is a standard value. The measured strength value
must be included between 159 and 161 A. If this value is out of
range, contact your Chopin agent.
Press a second time on the small arrow.
The first value is a numerical value. The second value is an
internal resistance measurement (low value). The value of this
resistance is a standard value. The measured strength value must
be included between 1.59 and 1.61 A. If this value is out of range,
contact your Chopin agent.
Pressing the big arrow.
Pressing the big arrow switches the 40 mA, which is the strength
needed by the probe to electrochemically produce iodine.
To verify this strength, remove the covers of the apparatus.
To be performed only by a CHOPIN representative.