Digital Power Meter 66200 Series User’s Manual
FETCh [ : SCALar ] : FREQuency?
MEASure [ : SCALar ] : FREQuency? 66201 | 66202
Description: These queries return the frequency in Hertz.
Setting Syntax:
none
Setting Parameters: none
Query Syntax: FETCh:FREQuency?<PMT> , MEASure:FREQuency?<PMT>
Response Parameters: <NR2>
Header on: :FETCH:FREQUENCY<space><NR2><RMT>
Header off: <NR2><RMT>
Example: none
FETCh [ : SCALar ] : POWer : REAL?
MEASure [ : SCALar ] : POWer : REAL? 66201 | 66202
Description:
These queries return the true power.
Setting Syntax: none
Setting Parameters: none
Query Syntax: FETCh:POWer:REAL?<PMT> , MEASure:POWer:REAL?<PMT>
Response Parameters: <NR2>
Header on: :FETCH:POWER:REAL<space><NR2><RMT>
Header off: <NR2><RMT>
Example: none
FETCh [ : SCALar ] : POWer : APParent?
MEASure [ : SCALar ] : POWer : APParent? 66201 | 66202
Description: These queries return the apparent power.
Setting Syntax:
none
Setting Parameters: none
Query Syntax: FETCh:POWer:APParent?<PMT> , MEASure:POWer:APParent?<PMT>
Response Parameters: <NR2>
Header on: :FETCH:POWER:APPARENT<space><NR2><RMT>
Header off: <NR2><RMT>
Example: none
FETCh [ : SCALar ] : POWer : REACtive?
MEASure [ : SCALar ] : POWer : REACtive? 66201 | 66202
Description: These queries return the reactive power.
Setting Syntax:
none
Setting Parameters: none
Query Syntax: FETCh:POWer:REACtive?<PMT> , MEASure:POWer:REACtive?<PMT>
Response Parameters: <NR2>
Header on: :FETCH:POWER:REACTIVE<space><NR2><RMT>
Header off: <NR2><RMT>
Example: none