16
NEAR FIELD GROUP
Number of targets: 9
Depth range: 2.0 - 10.0 mm
Spacing: 1.0 mm
VERTICAL DISTANCE GROUP
Number of targets: 17
Depth range: 1.0 - 18.0 cm
Spacing: 1.0 cm
a
HORIZONTAL DISTANCE: LINEAR PROBES
Number of targets: 5
Depth from Scan Surface #1: 5.0 cm
Spacing: 2.0 cm
b
HORIZONTAL DISTANCE: SECTOR SCANNING PROBES
Number of targets: 5
Depth from Scan Surface #1: 13.0 cm
Spacing: 2.0 cm
c
AXIAL / LATERAL RESOLUTION GROUPS
Number of Targets: 11
Lateral Displacement: 1.0 mm
Spacing: 4, 3, 2, 1, 0.5 & 0.25 mm
Depth: 7.0, 11.0, 4.0, 16.0 cm
CONTRAST TARGETS (CYSTS)
Type: Non-echogenic, cylindrical
Number of Groups: 5
Group 2, 3, 4:
Number of targets: 17
Diameter of targets: 2.0, 3.0, 4.0 mm
Depth of Targets: 1.0 - 17.0 cm
Spacing: 1.0 cm
Group 6, 8:
Number of targets: 8
Diameter of targets: 6.0, 8.0 mm
Depth of Targets: 1.0 - 17.0 cm
Spacing: 2.0 cm
a
Corrected distance. The physical distance between pins has been adjusted to account for the difference in
sound speed between urethane and tissue.
b
Uncorrected distance. Sound speed does not affect horizontal distance measurements when using a linear
probe
c
Corrected distance. Sound speed will affect horizontal distance measurements when using a sector scan-
ning probe