SLEEPVIEW USER MANUAL
10
Cleaning Supplies:
The following supplies are commonly used. They are recommended; however, similar
products of equal strength may be chosen.
Sani-Cloth
®
CaviWipes
®
Cavicide
®
Procedure:
1. Gather belts, pulse oximeter, re-usable nasal sensor/thermistor
2. Remove adhesive left over from any tape if tape was used
3. Wipe down thoroughly using disinfecting cloths or spray
4. Wipe dry with clean cloth,
5. Store item for next usage.
6. The exterior of the SleepView must also be wiped clean with a disinfecting cloth
or spray or a standard household surface cleanser.
Maintenance
SleepView does not require calibration. The SleepView software can be updated by end
customers. The SleepView Manager software will alert users that an update is available
as long as the computer is connected to the internet. Follow the software prompts to
download and install any updates. Please contact customer service with any questions.