SLEEPVIEW USER MANUAL
34
Recommended separation distances between portable and mobile RF communications equipment
and the SleepView system
The SleepView system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the SleepView system can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the SleepView system as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
π = 1.2 βπ
80 MHz to 800 MHz
π = 1.2 βπ
800 MHz to 2.5 GHz
π = 2.3 βπ
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
Guidance and manufacturerβs declaration β electromagnetic emissions
The SleepView System is intended for use in the electromagnetic environment specified below. The customer or the user of the
SleepView System should assure that it is used in such an environment.
Electromagnetic environment - guidance
The SleepView uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
The SleepView System is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3