EasyManua.ls Logo

Clevemed SleepView - RF Emissions and Separation Distances

Clevemed SleepView
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SLEEPVIEW USER MANUAL
34
Recommended separation distances between portable and mobile RF communications equipment
and the SleepView system
The SleepView system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the SleepView system can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the SleepView system as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
𝑑 = 1.2 βˆšπ‘ƒ
80 MHz to 800 MHz
𝑑 = 1.2 βˆšπ‘ƒ
800 MHz to 2.5 GHz
𝑑 = 2.3 βˆšπ‘ƒ
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
Guidance and manufacturer’s declaration – electromagnetic emissions
The SleepView System is intended for use in the electromagnetic environment specified below. The customer or the user of the
SleepView System should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The SleepView uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The SleepView System is suitable for use in all establishments, including
domestic establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Not applicable

Other manuals for Clevemed SleepView