2. Safety notes
Coolant
The Osscora surgical set is designed for use with
physiological saline solution. Use only suitable irrigation
fluids and comply with the medical data and instructions
from the manufacturer. Use the supplied irrigation tubing
set or accessories supplied by Cochlear. Coolant bottle or
coolant bag can be obtained from a pharmacist.
Sterility of irrigation tubing set
Sterile irrigation tubing sets are supplied with the
equipment. These irrigation tubings are disposable articles
and must be thrown away after each treatment! Please
note the expiry date and the relevant regulations for
disposal of irrigation tubing. Only use disposable irrigation
tubing with undamaged packaging.
Rotational energy
Fast deceleration of the bur can, at times, cause the
selected torque to be overloaded as a result of the
rotational energy stored in the drive system.
Perform a test run prior to thread cutting and implant
placement to check if motor stops correctly after
releasing the foot control.
Mains cable
Only use the mains cable supplied.
Only connect to an earthed socket outlet.
Set up the device so that the power switch is easily
accessible. In dangerous situations, the device can be
disconnected from the power supply using the power
switch or power cable. The power switch can also be used
to safely stop the device.
Power failure
In the event of a power failure, if the control unit is
switched off, or when alternating between programs, the
last values set are saved and reactivated on power-up.
System failure
A total system failure of the Osscora surgical set does not
constitute a critical fault. Simply switch the unit off and
then on again.
Intermittent operating mode S3 (3min/10min)
If the operating mode specified is observed no overheating
of the system and therefore no injury to the patient, user or
third persons arises.
The responsibility for the use and timely shutdown of the
system lies with the user.