16. Servicing
Regular checking of the foot control
Visual inspection for outside damage.
Measurement of enclosure leakage current.
Functional test with check if the maximum speed can
be reached.
- Measurement of ESD conductivity of the foot control
(electro-static discharge)
We recommend that only authorised personnel undertake this
servicing and checking. For details, please contact your local
Cochlear office.
Regular checking of Osscora surgical set and
accessories
Regular servicing of function and safety of the
Osscora surgical set including the accessories
is necessary and should be carried out at least
once every three years, unless shorter intervals
are prescribed by law. The inspection must be
undertaken by a qualified organisation and needs
to include the following procedures:
Visual inspection for outside damage.
Measurement of device leakage current.
Measurement of patient leakage current.
Visual inspection of internal components and electronic
boards on suspicion of safety interference, e. g. mechanical
damage of the enclosure or indicators for overheated or
burned (electronic) components.