50
The foot control is watertight according to IPX8, 1m depth of
immersion, 1 hour (water-tight according to IEC 60529).
Pollution degree: 2
Overvoltage category: II
Altitude: up to 3,000 m above sea level
17. Technical data
Classification according to § 5 of the General Specifications
for the Safety of Medical Electrical Equipment according to
IEC 60601-1/UL 60601-1:
Equipment of Protective Class II (type of protection against
electric shock).
Type B appliance (not suitable for intracardiac application).
Foot control conforms to Class AP according
to IEC 60601-1/UL 60601-1 in danger zone M.
91053
Supply voltage: 100 – 130 V
Permitted voltage fluctuation: ± 10 %
Nominal current: 0.2 – 1.7 A
Frequency: 50 – 60 Hz
Mains fuse: 2x 250 V – T1.25AH
Max. power consumption: 170 VA
Max. mechanical output power: 70 W
Max. torque on the motor: 5.5 Ncm
Motor speed range in
the nominal voltage range: 300 – 40,000 min.-1
Max. coolant in ml: approx. 100 ml/min
Operating mode: S3 (3min / 10min)
Dimensions in mm (WxDxH): 252 x 254 x 90
Set weight in kg: 4.10