16. Servicing
Motor and cable
The standard ISO 11498 stipulates a durability of at least 250
sterilisation cycles. In the case of the motor with cable from
Cochlear we recommend that you carry out a regular servicing
after 500 sterilisation cycles, or one year.
Repairs
If a defect occurs, always return all the equipment, due to
the fact that with motor malfunctions, an inspection of the
electronic controls is also necessary!
Returns
Refer all questions to your local Cochlear office.
Always return equipment in the original packaging!
Do not coil the cable around the motor and do not
twist or kink the motor cable! (Risk of damage).