Cochlear Baha - 12. Error Messages; Troubleshooting Common Error Codes

Cochlear Baha
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
English
27
Error No. Description Remedy
00 Electronics overheating–safety shutdown Switch off the equipment, allow to cool for at least
10 minutes, then restart
01 Electronics overloaded Switch off the equipment, allow to cool for at least
10 minutes, then restart
07 Foot control error–initialising Switch off the equipment, restart, do not actuate
foot control when switching on
09 Foot control error Switch off the equipment, check connection of the
foot control, then restart
19 Run limiting control Switch off the equipment and restart
99 System failure Switch off the equipment, allow to cool for at least
10 minutes, then restart
Motor temperature too high
Motor connection – safety shutdown
Switch off the equipment, check motor connection,
allow motor to cool for at least 10 minutes,
then restart
12. Error messages