31
The front panel of the control unit and the foot
control are sealed and may be wiped clean.
– Disinfection with surface disinfectants, wiping
disinfection is recommended.
– Use only surface disinfectants which do not contain
chlorine and which are certified by officially
recognised institutes.
For USA: Use EPA registered surface disinfectants.
– Note the manufacturer’s specifications for the use of
the surface disinfectants.
– Clean the ESD spring contact of the foot control
regularly.
13. Reprocessing guidelines
Manual cleaning and disinfection of the
motor with cable
Do not twist and kink the motor cable! Do not coil it
too tightly!
Before disinfection and cleaning push the motor
protective cover onto the motor attachment!
Pre-disinfection
– If heavily soiled: Clean first with disinfectant cloths.
Only use disinfectants that have no protein-fixing
effects.
Manual cleaning
– Rinse under demineralised water (< 38 °C) with the aid of a
brush (brush is not recommended in U.K.)
– Then remove any liquid residues (absorbent cloth, blow dry
with compressed air).
– Do not place the motor in liquid disinfectant or
in an ultrasonic bath.
Disinfection
– Disinfect with surface disinfectants. Wiping disinfection
is recommended.
– Use only surface disinfectants which do not contain chlorine
and which are certified by officially recognised institutes.
For USA: Use EPA registered surface disinfectants.
– Note the manufacturer’s specifications for the use of the
surface disinfectants.