EasyManua.ls Logo

Cochlear Nucleus 8 - Environmental Conditions and Battery Use

Cochlear Nucleus 8
120 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Other information
106 Nucleus 8 Sound Processor User Guide
Environmental conditions
Condition Minimum Maximum
Storage and transport
temperature
-10° C (+14° F) +55° C (+131° F)
Storage and transport humidity 0% RH 90% RH
Operating temperature
(sound processor)
+5° C (+41° F) +40° C (+104° F)
Operating temperature
(battery charger)
0° C (+32° F) +40° C (+104° F)
Operating relative humidity 0% RH 90% RH
Operating pressure 700 hPa 1060 hPa
The temperature of the sound processor and its accessories may
rise by up to 2.7° C (4.86° F) during normal operation, and could
result in these components reaching a temperature of
+42.7° C (+108.86° F) when operated at the maximum environmental
temperature of +40.0° C (+104° F).
Disposable batteries
Check the battery manufacturer’s recommended operating
conditions for disposable batteries used in your sound
processor.