Troubleshoot
86 Nucleus 8 Sound Processor User Guide
Battery
Problem Resolution
Batteries are not
lasting as long as
usual
1� Clean all connections and parts of the
battery module.
2� Clean all connections on the
processing unit.
3� Make sure you are using the correct
coil magnet for your implant. If unsure,
contact your clinician.
4� Try replacing the coil with a new coil.
5� If you are using disposable batteries,
check that you are using the
recommended batteries. SeeChange
disposable batteries on page 21.
6� Don’t forget to let new disposable
batteries stand for a few seconds
before you put them in the processing
unit.
7� If the problem continues, contact your
clinician.
The indicator light
on the battery
charger shows a fully
charged battery is
still charging
This won’t cause any damage to the
battery as the charge cycle on a fully
charged battery is very short.