1. Select correct tubing size and flow rate.
2. Press CAL, calibration volume appears.
3. Press STOP/START, the pump will use its stored memory to dispense
the specified calibration sample quantity. The pump will stop
automatically.
4. Weigh/measure the sample.
5. Use UP/DOWN arrow keys to correct the flashing display.
NOTE: If the adjusted calibration is too great, “Err” will appear in the
display. If this occurs, press the CAL control and repeat the calibration
procedure. The microprocessor will retain one special calibration value per
tubing size, even when power is turned off. The next calibration will
replace the existing value.
6. Press size to exit the calibration cycle.
1. To set the maximum flowrate for non-standard pumps or tubing sizes,
OTHER press CAL, then FLOW. The maximum flowrate will then
flash on the display.
2. Use UP/DOWN arrow keys to set desired flowrate.
3. Press SIZE to exit.
B/T
®
RAPID-LOAD
®
Peristaltic Pumps and Drive Operating Manual 3-3Cole-Parmer
Section 3
Operation
Maximum Flowrate
(OTHER Tubing)
Calibration

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Cole Parmer Masterflex B/T Rapid-Load and is the answer not in the manual?

Cole Parmer Masterflex B/T Rapid-Load Specifications

General IconGeneral
BrandCole Parmer
ModelMasterflex B/T Rapid-Load
CategoryWater Pump
LanguageEnglish

Related product manuals