NOTE: The maximum flow rate for a tubing will change after a calibration
is performed. To retain control of the entire flow range, the “HI” scale
setting must be changed to the new maximum flow rate after a calibration
is performed.
6. Remote STOP/START can be configured to be optional (“OFF”) or
mandatory (“ON”). When “ON” is selected, drive will not run unless
remote STOP/START is closed When “OFF” is selected (factory
default), remote STOP/START can be used to start drive, but drive
can also be started by keypad or remote inputs when remote STOP/
START is open. Internal mode or remote mode (mA or V) each have
their own STOP/START configuration, so first select the desired
operating mode before changing STOP/START setup.
a) Press and hold the MODE key until the display changes to
“STOP”. The display will alternate with an “ON” or “OFF”.
b) Use the UP/DOWN arrow keys to select “ON” or “OFF”.
c) Press any other key to save and exit.
3-6 B/T
®
RAPID-LOAD
®
Peristaltic Pumps and Drive Operating Manual Cole-Parmer
Section 3
Operation
Remote Control Setup
(continued)
Figure 3-3. Remote Control Connector Pin Configuration