4.5.9 Partial buttons or all buttons failure
Partial buttons or all
buttons failure
Connector between the control
panel and the line board out of
proper connection.
open the upper cover of the machine and
re-connect the connector between control
panel and line board
Internal line between membrane
buttons is broken due to long time
of use.
Replace the involved membrane buttons.
Change the new main board.
4.5.10 Print problem.
Waveforms displayed
properly but can’t be
printed
Print paper in incorrect position
Load print paper in correct position
Check whether the print paper in the right
position.
Print cable didn’t connect well
Check the print cable is reliably connected