Shenzhen Comen Medical Instruments co., ltd.
21
50mm/s,error limit±2%.
5.2.2.2 Test method
a)Set the sensitivity of electrocardiograph as 10 mm/mv,print speed 25mm/s, AC filter 50Hz,
EMG filter off,low-pass filter 150Hz,drift filter 0.05Hz,manual model.Switch on the ST light of the
detector,press S+SU,keep the indicator of detector“3”lit,Change the output value 20μ V、10Hz
to 0.5mV、10Hz.
b)Enter a triangular-wave signal of frequency of 25Hz±1%、0.5mVp-p,set the print speed as
25mm/s ,after paper advanced 1s,measure by square master.
5 Groups continuous sequence(each group with 10 cycles), the distance of each sequence on
the print paper is 10±0.5mm,the distance of 50 cycles on the print paper should be 50±2.5mm.
c)Set the print speed as 5mm/s,and set the signal frequency as 5Hz±1%,repeated test.
d)Set the print speed as10mm/s,and set the signal frequency as 10Hz±1%,repeated test.
e) Set the print speed as 12.5mm/s,and set the signal frequency as 12.5Hz±1%,repeated test.
f) Set the print speed as 50mm/s,turn off the AC filter,and set the signal frequency as 50Hz±1%,
repeated test.