Shenzhen Comen Medical Instruments co., ltd.
3
application.
WARNING :
♦ Only patient cable and other accessories supplied by our company can
be used. Or else, the performance and electric shock protection can not
be guaranteed.
♦ Be sure that all electrodes have been connected to the patient correctly
before operation.
♦ Be sure that the conductive parts of electrodes and associated
connectors, including neutral electrode, should not contact with earth
or any other conducting objects.
♦ Electrodes with defibrillator protection should be used while
defibrillating.
♦ There is no danger for patients with pacemaker.
♦ Do not touch the patient, bed, table and the equipment while using
defibrillator or pacemaker simultaneously.
♦ In order to avoid burning, please keep the electrode far away from the
radio knife while using electrosurgical equipment simultaneously.
WARNING :
♦ Accessory equipment connected to the analog and digital interfaces
must be certified according to the respective IEC standards (e.g. IEC
60950 for data processing equipment and IEC 60601-1 for medical
equipment). Furthermore all configurations shall comply with the valid
version of the standard IEC 60601-1-1. Therefore anybody, who
connects additional equipment to the signal input connector or output
connector to configure a medical system, must make sure that it
complies with the requirements of the valid version of the system
standard IEC 60601-1-1. If in doubt, consult our technical service
department or your local distributor.
♦ The summation of leakage current should never exceed leakage current
limits while several other units are used at the same time.
♦ The potential equalization conductor can be connected to that of other
equipment when necessary, to make sure that all these equipment are