8. EMC TABLE
The Compex stimulator needs special precautions regarding EMC and needs to be installed and put into service according to the
EMC information provided.
Portable and mobile RF communications equipment can aect the Compex stimulator.
The use of Accessories, transducers, and cables other than those specied by the manufacturer, may result in increased Emissions
or decreased Immunity of the Compex stimulator.
The Compex stimulator should not be used adjacent to or stacked with other equipment and if adjacent or stacked use is necessary,
the Compex stimulator should be observed to verify normal operation in the conguration in which it will be used.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The Compex stimulator is intended for use in the electromagnetic environment specied below. The customer or the user of the Compex
stimulator should assure that it is used in such an environment.
EMISSION TESTS COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT
– GUIDANCE
Radiated Emissions CISPR 11 Group 1
The Compex uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
Conducted Emissions CISPR 11 Class B
Harmonic emissions IEC 61000-3-2 Class A
The Compex is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Voltage uctuations IEC 61000-3-3 Complies