IEC 60601-1-1. Everybody who connects additional equipment to the signal
input connector or signal output connector configures a medical system, and is
therefore responsible that the system complies with the requirements of the
valid version of the system standard IEC 60601-1-1. If in doubt, consult our
technical service department or your local distributor.
CAUTION : The device is designed for continuous and is “ordinary” (i.e. not drip or
splash-proof).
CAUTION : Keep the environment clean. Avoid vibration. Keep it far from corrosive
medicine, dust area, high-temperature and humid environment.
CAUTION : Do not operate the unit if it is damp or wet because of condensation or spills.
Avoid using the equipment immediately after moving it from a cold
environment to a warm, humid location.
CAUTION : Do not immerse transducers in liquid. When using solutions, use sterile wipes
to avoid pouring fluids directly on the transducers.
CAUTION
④
Do not use high temperature heating or gas to disinfect the monitor and its
accessories.
CAUTION
④
Turn off the power supply before clean the machine.
CAUTION : The temperature should not exceed 6
0℃
when clean
the belt.
CAUTION : Electromagnetic Interference-Ensure that the environment in which the
fetal monitor is installed is not subject to any sources of strong
electromagnetic interference, such as radio transmitters, mobile telephones,
etc.
CAUTION : The monitor must serviced by proper training and knowledge, practical
personnel. The recommended testing interval is once twice year or under the
leakage current measurement and insulation testing.