ECG1200G Electrocardiograph - User Manual
I
Statement
Our company owns all rights to this unpublished work and intends to maintain it as confidential
information. This publication is used only for reference of operation, maintenance, or repair of
our device. No part of this can be disseminated to others.
This document contains proprietary information, which is protected by copyright. All rights
reserved. Photocopy, reproduction or translation of any part in the manual without our company's
written permission is prohibited.
All information contained in this publication is believed to be correct. Our company shall not be
liable for incidental and consequential damages in connection with the furnishing, performance,
or use of this material. This publication may refer to information and protected by copyrights or
patents and does not convey any license under the patent rights of our company, nor the rights of
others.Our company does not assume any liability for arising out of any infringements of patents
or other rights of the third parties.
Content of this manual is subject to change without prior notice.

Questions and Answers

  • B
    Benjamin HernandezJul 27, 2025
    What to do if Contec Medical Equipment shows too large interference, disorderly waveform?
    • M
      moorecrystalJul 27, 2025
      If you are experiencing large interference and a disorderly waveform, check the lead cables, grounding cable, and power connections to ensure they are reliably connected. Also, if the patient is nervous and cannot keep quiet, try to treat the patient well to calm them.
  • A
    Amanda WheelerAug 3, 2025
    What to do if Contec Medical Equipment shows not regular waveform, large up-and-down, beeline figure?
    • L
      laura88Aug 3, 2025
      If the waveform is not regular and shows large up-and-down movements or a beeline figure, try the following: Use high-quality alcohol to improve electrode conductivity. Charge the battery if the capacity is low. Clean the electrode slices and the skin below them with alcohol to improve the connection. Ensure that the lead cables are firmly connected to the device's plug, and that the electrodes are properly connected to the lead cables.
  • G
    Garrett JacksonAug 8, 2025
    Why does my Contec Medical Equipment have a baseline burr?
    • C
      Caitlin CohenAug 8, 2025
      A baseline burr may be due to large AC interference or the patient being nervous, leading to large EMG interference; in both cases, try to amend the environment. Also, if the bed is made of steel, replace it. Ensure that the power cable and lead cables are not parallel or too close to each other.
  • F
    Frank BentonAug 14, 2025
    What to do if Contec Medical Equipment has a baseline draft?
    • T
      Taylor GregoryAug 14, 2025
      If there's a baseline draft, it could be due to low power; in this case, charge the battery. Alternatively, the patient's movement could be the cause, so ensure the patient remains still.
  • E
    Evan OchoaAug 20, 2025
    How to fix unclear waveform on Contec Medical Equipment?
    • L
      Lisa SullivanAug 20, 2025
      If the waveform is unclear, the printer head surface may be dirty. If there's a power cut, clean the printer head with alcohol and wait until it is dry before printing. Alternatively, the thermal paper could be the problem, so replace it with the specified thermal print paper.

Summary

ECG1200 G Overview

Key Technical Specifications

Lists key operational and technical parameters of the electrocardiograph.

Safety Precautions

ECG Measurement Analysis Notes

Details limitations and considerations for accurate ECG waveform measurement and analysis.

Product Main Characteristics

Pre-Operation Guidelines

Operation Preparation Steps

Operational Guidance

Recording Paper Handling

Electrode Placement Instructions

Grounding and Power Connection

Battery Operation and Status

Control Panel and Interface Navigation

Main Interface Layout

Introduces the main interface, its layout, and key function buttons for device operation.

ECG Sampling Interface

Guides on entering and operating the sampling interface, including lead display and stop functions.

Patient Case Information Input

Instructions on how to input patient case information using the device's interface.

Patient Case Management

Details managing patient case data, including viewing, editing, and deleting records.

ECG Sampling Configuration

Configuration options for ECG sampling parameters, including filters, lead display, and speed.

Troubleshooting Common Issues

Troubleshooting Solutions List

Device Maintenance Procedures

Contec ECG1200G Specifications

General IconGeneral
Type12-channel ECG
Enduring Polarization Voltage±300mV
Noise Level≤15μVp-p
Battery Life≥2 hours
Input Impedance≥50MΩ
Lead12 leads
DisplayLCD
Frequency Response0.05Hz - 150Hz
Time Constant≥3.2s
Input Circuit Current≤0.1μA
Sensitivity5, 10, 20, 50mm/mV
Standard Sensitivity10mm/mV
Recording WayThermal printing system
Paper Speed5, 10, 12.5, 25, 50mm/s
Record ModesAuto, Manual, Rhythm
Power SupplyAC 100-240V, 50/60Hz
ConnectivityUSB
Bandwidth0.05Hz - 150Hz