ECG1200G Electrocardiograph - User Manual
1
Chapter1 Overview
1.1 Overview
12-channel Electrocardiograph is such a kind of electrocardiograph, which gathers 12-lead ECG
signals simultaneously and prints out the ECG waveforms with thermal printing system. It's
major functions are as follows: recording and displaying ECG waveforms in manual/auto ways;
measuring and diagnosing ECG waveform parameters automatically; electrode-off and paper
lack prompt; optional interface languages (Chinese/English); powered either by AC or DC;
selectable rhythm lead to detect abnormal heart rate conveniently ;case database management,
etc.
800*600 dots high resolution LCD screen; 1728 dot wide, it can print 12-channel ECG
waveforms. You can operate either by buttons or touch screen, which is convenient and quick.
1.2 Main Technical Specifications
1.2.1 Environment conditions
Operation
a) Environment temperature: +5~+35
b) Relative humidity: 80%
c) Power supply: AC:100V~240V,50/60Hz
DC:14.8V, 3700 mAh rechargeable lithium battery
d) Atmospheric pressure: 86kPa~106kPa
Transportation and Storage
a) Environment temperature: -10~55
b) Relative humidity: 95%
c) Atmospheric pressure: 50kPa~106kPa
1.2.2 Input way: Floating and defibrillation protection
1.2.3 Lead: Standard 12 leads(model:BIP0057)
1.2.4 Patient leak current: <10µA
1.2.5 Input impedance: 50M
1.2.6 Frequency response: 0.05Hz~150Hz(-3dB~+0.4dB)
1.2.7 Time constant: Time constant3.2s
1.2.8 CMRR: >60dB, >100dB( Adding filter)
1.2.9 EMG interference filter: 25Hz/35Hz(-3dB)
1.2.10 Recording way: Thermal printing system
1.2.11 Specification of recording paper: 210mm(W)*20m(L) high-speed thermal paper
1.2.12 Paper speed:
Auto record: 25mm/s, 50mm/s, error:±5%
Rhythm record: 25mm/s, 50mm/s, error:±5%
Manual record:5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s, error:±5%
1.2.13 Sensitivity selections: 2.5,5,10,20,40mm/mV, error:±5%. Standard sensitivity is
10mm/mV±0.2mm/mV.
1.2.14 Auto record: Record setup according to auto record format and mode, automatically
changing leads, measuring and analysing.
1.2.15 Rhythm record: record setup according to rhythm record format and mode, automatically
measuring and analysing.
1.2.16 Manual record: record setup according to manual record format, manually changing leads.
1.2.17 Measurement parameters: HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T
interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5)+S(V1).
1.2.18 Product safety type: Class I CF applied part; there is defibrillation and pacing protection
circuit in it.
1.2.19 Enduring polarization voltage: ±500mV
1.2.20 Noise level: 15µVp-p
1.2.21 Fuse specification: 2 pcs φ5*20mm AC time lag; T1.6A/250V(Power supply 220V)
1.2.22 Size: 340mm(L)*320mm(W)*85mm(H)