63
TECHNICAL
& CARE
CAUTION: Avoid exposing meter and test strips to excessive humidity, heat, cold,
dust,or dirt.
Caring for Your Meter
Caring for Your Meter
Store your meter in the carrying case provided wheneverpossible.
Wash hands and dry well before handling to keep the meter and test strips free of
water, oils and othercontaminants.
Keep blood, control solution, and cleaning fluid from entering the test strip port or USB plug.
Handle the meter carefully to avoid damaging the electronics or causing other malfunctions.
The USB cap of your C
ONTOUR
®
NEXT LINK 2.4 meter is designed to protect the USB plug.
Should you lose the cap, contact Customer Service for a replacement. Please see back
cover for contact information.
Cleaning and Disinfection
Your CONTOUR
®
NEXT LINK 2.4 meter should be cleaned and disinfected once a week.
For the meter, use only Clorox
®
Germicidal Wipes containing 0.55% sodium hypochlorite
(bleach), which has been proven to be safe to use with the C
ONTOUR NEXT LINK 2.4 meter
and lancing device.
Cleaning is the removal of visible dirt and debris, but does not reduce the risk for
transmission of infectious diseases. Your C
ONTOUR NEXT LINK 2.4 system should be
cleaned of dirt and debris once a week.
Disinfecting (if performed properly) reduces the risk of transmitting infectious diseases.
Your meter and lancing device should be disinfected once a week.
WARNING
Always wash your hands well with soap and water before and after testing
and handling the meter, lancing device, or test strips.
NOTE: If your meter is being operated by a second person who is providing testing
assistance to you, the meter and lancing device should be disinfected prior to use
by the second person.
Signs of Deterioration
The cleaning and disinfecting directions provided should not cause any damage or
degradation to the external case, buttons or display. Your CONTOUR NEXT LINK 2.4 meter
has been tested for 260 cycles of cleaning and disinfection (one cycle per week for 5 years).
This device has been demonstrated to withstand 5 years of cleaning and disinfection
without damage. You should contact Customer Service for assistance if your device
malfunctions for any cause or if you notice any changes in the meter case or display.
Clorox Germicidal Wipes are available for purchase online at http://www.officedepot.com
or http://www.amazon.com or for more information call Ascensia Diabetes Care Customer
Service. Contact information for Customer Service is on the back cover.
For more information see:
“FDA Public Health Notification: Use of Fingerstick Devices on More than One Person
Poses Risk for Transmitting Bloodborne Pathogens: Initial Communication” (2010).
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm
“CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk
for Transmitting Bloodborne Pathogens” (2010).
http://www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html
85587269_CntrNextLINK_2_4_UG_EN_FpBp_v4.indd 63 11/18/16 9:03 AM
TECHNICAL
& CARE
62
System Specifications
Test Sample: Capillary whole blood
Test Result: Referenced to plasma/serum glucose
Sample Volume: 0.6 μL
Measuring Range: 20 mg/dL – 600 mg/dL
Countdown Time: 5 seconds
Memory: Stores most recent 1000 test results
Battery Type: Non-serviceable, 250mAh rechargeable lithium polymer battery, 3.4V –
4.2V (5V input voltage)
Meter/Battery Life: 5 years
Charging Current: 300 mA
Strip Storage Temperature Range: 41ºF – 86ºF
Control Storage Temperature Range: 48ºF – 86ºF
Meter Operating Temperature Range: 41ºF – 113ºF
Control Solution Operating Temperature Range: 59°F–95°F.
Meter Operating Humidity Range: 10%–93% RH
Dimensions: 3.8 in wide x 2.1 in high x 0.7 in thick
Weight: 43 grams
Sound Output: 45 to 80 dBA at a distance of 3.9 in
Radio Frequency Communication: 2.4 GHz
Electromagnetic Compatibility (EMC): The C
ONTOUR
®
NEXT LINK 2.4 meter complies
with the electromagnetic requirements specified in ISO 15197: 2013. Electromagnetic
emissions are low and unlikely to interfere with other nearby electronic equipment,
nor are emissions from nearby electronic equipment likely to interfere with the
C
ONTOUR NEXT LINK 2.4. The CONTOUR NEXT LINK 2.4 meter meets the requirements
of IEC 61326-2-6 for immunity to electrostatic discharge. It is advisable to avoid use
of electronic devices in very dry environments especially if synthetic materials are
present. The CONTOUR NEXT LINK 2.4 meter meets the requirements of IEC 61326-1
for radio frequency interference. To avoid radio frequency interference do not use the
CONTOUR NEXT LINK 2.4 meter near cellular or cordless telephones, walkie talkies,
garage door openers, radio transmitters or other electrical or electronic equipment
that are sources of electromagnetic radiation, as these may interfere with the proper
operation ofthe meter.
®
System Specifications
85587269_CntrNextLINK_2_4_UG_EN_FpBp_v4.indd 62 11/18/16 9:03 AM
63
TECHNICAL
& CARE
CAUTION: Avoid exposing meter and test strips to excessive humidity, heat, cold,
dust,or dirt.
Caring for Your Meter
Caring for Your Meter
Store your meter in the carrying case provided wheneverpossible.
Wash hands and dry well before handling to keep the meter and test strips free of
water, oils and othercontaminants.
Keep blood, control solution, and cleaning fluid from entering the test strip port or USB plug.
Handle the meter carefully to avoid damaging the electronics or causing other malfunctions.
The USB cap of your C
ONTOUR
®
NEXT LINK 2.4 meter is designed to protect the USB plug.
Should you lose the cap, contact Customer Service for a replacement. Please see back
cover for contact information.
Cleaning and Disinfection
Your CONTOUR
®
NEXT LINK 2.4 meter should be cleaned and disinfected once a week.
For the meter, use only Clorox
®
Germicidal Wipes containing 0.55% sodium hypochlorite
(bleach), which has been proven to be safe to use with the C
ONTOUR NEXT LINK 2.4 meter
and lancing device.
Cleaning is the removal of visible dirt and debris, but does not reduce the risk for
transmission of infectious diseases. Your C
ONTOUR NEXT LINK 2.4 system should be
cleaned of dirt and debris once a week.
Disinfecting (if performed properly) reduces the risk of transmitting infectious diseases.
Your meter and lancing device should be disinfected once a week.
WARNING
Always wash your hands well with soap and water before and after testing
and handling the meter, lancing device, or test strips.
NOTE: If your meter is being operated by a second person who is providing testing
assistance to you, the meter and lancing device should be disinfected prior to use
by the second person.
Signs of Deterioration
The cleaning and disinfecting directions provided should not cause any damage or
degradation to the external case, buttons or display. Your CONTOUR NEXT LINK 2.4 meter
has been tested for 260 cycles of cleaning and disinfection (one cycle per week for 5 years).
This device has been demonstrated to withstand 5 years of cleaning and disinfection
without damage. You should contact Customer Service for assistance if your device
malfunctions for any cause or if you notice any changes in the meter case or display.
Clorox Germicidal Wipes are available for purchase online at http://www.officedepot.com
or http://www.amazon.com or for more information call Ascensia Diabetes Care Customer
Service. Contact information for Customer Service is on the back cover.
For more information see:
“FDA Public Health Notification: Use of Fingerstick Devices on More than One Person
Poses Risk for Transmitting Bloodborne Pathogens: Initial Communication” (2010).
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm
“CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk
for Transmitting Bloodborne Pathogens” (2010).
http://www.cdc.gov/injectionsafety/Fingerstick-DevicesBGM.html
85587269_CntrNextLINK_2_4_UG_EN_FpBp_v4.indd 63 11/18/16 9:03 AM
TECHNICAL
& CARE
62
System Specifications
Test Sample: Capillary whole blood
Test Result: Referenced to plasma/serum glucose
Sample Volume: 0.6 μL
Measuring Range: 20 mg/dL 600 mg/dL
Countdown Time: 5 seconds
Memory: Stores most recent 1000 test results
Battery Type: Non-serviceable, 250mAh rechargeable lithium polymer battery, 3.4V
4.2V (5V input voltage)
Meter/Battery Life: 5 years
Charging Current: 300 mA
Strip Storage Temperature Range: 41ºF 86ºF
Control Storage Temperature Range: 48ºF 86ºF
Meter Operating Temperature Range: 41ºF 113ºF
Control Solution Operating Temperature Range: 59°F–95°F.
Meter Operating Humidity Range: 10%–93% RH
Dimensions: 3.8 in wide x 2.1 in high x 0.7 in thick
Weight: 43 grams
Sound Output: 45 to 80 dBA at a distance of 3.9 in
Radio Frequency Communication: 2.4 GHz
Electromagnetic Compatibility (EMC): The C
ONTOUR
®
NEXT LINK 2.4 meter complies
with the electromagnetic requirements specified in ISO 15197: 2013. Electromagnetic
emissions are low and unlikely to interfere with other nearby electronic equipment,
nor are emissions from nearby electronic equipment likely to interfere with the
C
ONTOUR NEXT LINK 2.4. The CONTOUR NEXT LINK 2.4 meter meets the requirements
of IEC 61326-2-6 for immunity to electrostatic discharge. It is advisable to avoid use
of electronic devices in very dry environments especially if synthetic materials are
present. The CONTOUR NEXT LINK 2.4 meter meets the requirements of IEC 61326-1
for radio frequency interference. To avoid radio frequency interference do not use the
CONTOUR NEXT LINK 2.4 meter near cellular or cordless telephones, walkie talkies,
garage door openers, radio transmitters or other electrical or electronic equipment
that are sources of electromagnetic radiation, as these may interfere with the proper
operation ofthe meter.
®
System Specifications
85587269_CntrNextLINK_2_4_UG_EN_FpBp_v4.indd 62 11/18/16 9:03 AM
85587269_CntrNextLINK_2_4_UG_EN_FpBp_v5_UnMaskedPlaced.pdf:66