EasyManua.ls Logo

Crown Micro-Tech 1200 Service Manual

Crown Micro-Tech 1200
4 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
Micro-Tech 1200 Amplifier Service Manual
13
Electrical Checkout Procedures
GENERAL INFORMATION
The following test procedures are to be used to verify
operation of this amplifier. DO NOT connect a load or
inject a signal unless directed to do so by the proce-
dure. These tests, though meant for verification and
alignment of the amplifier, may also be very helpful in
troubleshooting. For best results, tests should be
performed in order.
All tests assume that AC power is from a regulated 120
VAC source. Test equipment includes an oscilloscope,
a DMM, a signal generator, loads, and I.M.D. and
T.H.D. noise test equipment.
STANDARD INITIAL CONDITIONS
Level controls fully clockwise.
Stereo/Mono switch in Stereo.
Sensitivity switch in 26 dB fixed gain position.
It is assumed, in each step, that conditions of the
amplifier are per these initial conditions unless other-
wise specified.
TEST 1: DC OFFSET
Spec: 0 VDC, ±10 mV.
Initial Conditions: Controls per standard, inputs shorted.
Procedure: Measure DC voltage at the output connec-
tors (rear panel). There is no adjustment for output
offset. If spec is not met, there is an electrical malfunc-
tion. Slightly out of spec measurement is usually due to
U104/U204 out of tolorance.
TEST 2: OUTPUT BIAS ADJUSTMENT
Spec: 300 to 320 mVDC.
Initial Conditions: Controls per standard, heatsink tem-
perature less than 40°C.
Procedure: Measure DC voltages on the output module
across R02, adjust R26 if necessary. Measure DC
voltages on the output module across R21, adjust R23
if necessary. Repeat for second channel.
TEST 3: ODEP VOLTAGE ADJUSTMENT
Spec: Bias Per Chart, ±0.1V DC.
Initial Conditions: Controls per standard, heatsink at
room temperature 20 to 30°C (68 to 86°F). Note: This
adjustment should normally be performed within 2
minutes of turn on from ambient (cold) conditions. If
possible measure heatsink temperature, if not measure
ambient room temperature. Use this information when
referencing the following chart. The following is a list of
ODEP bias voltages VS. temperature.
°F °C V
–ODEP
V
+ODEP
66 18.9 –10.31 11.41
68 20.0 –10.26 11.36
70 21.1 –10.20 11.30
72 22.2 –10.14 11.24
74 23.3 –10.09 11.19
76 24.4 –10.03 11.13
77 25.0 –10.00 11.10
78 25.6 –9.97 11.07
80 26.7 –9.91 11.01
82 27.8 –9.86 10.96
84 28.9 –9.80 10.90
86 30.0 –9.74 10.84
88 31.1 –9.69 10.79
90 32.2 –9.63 10.73
92 33.3 –9.57 10.67
94 34.4 –9.51 10.61
–ODEP Procedure: Measure pin 6 of U100 and, if neces-
sary, adjust R121 to obtain V
–ODEP
as specified above.
Measure pin 6 of U200 and, if necessary, adjust R221
to obtain V
–ODEP
as specified above.
+ODEP Procedure: Measure pin 6 of U103 and, if neces-
sary, adjust R132 to obtain V
+ODEP
as specified above.
Measure pin 6 of U203 and, if necessary, adjust R232
to obtain V
+ODEP
as specified above.

Other manuals for Crown Micro-Tech 1200

Questions and Answers:

Crown Micro-Tech 1200 Specifications

General IconGeneral
Power Output per Channel (2 ohms)600W
THD< 0.05%
Input Impedance20k ohms balanced, 10k ohms unbalanced
Bridged Mono Output (4 ohms)1200W
Frequency Response20Hz - 20kHz
Input Sensitivity1.4V
Voltage Gain26dB
Dimensions19" x 3.5"
Signal to Noise Ratio>100dB
CoolingFan-cooled

Summary

General Information and Initial Conditions

Standard Initial Conditions

Defines the standard conditions for amplifier testing, including control settings and assumptions.

Electrical Checkout Procedures

DC Offset Test

Measures DC voltage at the output connectors to check for offset.

Output Bias Adjustment

Measures and adjusts DC bias voltages on the output module.

ODEP Voltage Adjustment

Adjusts ODEP bias voltage based on a temperature-dependent chart.

AC Power Draw Test

Measures maximum quiescent AC line wattage draw with no load or signal.

Common Mode Rejection Test

Tests the amplifier's ability to reject common mode signals at specified frequencies.

Voltage Gain Test

Verifies voltage gain across different sensitivity switch positions.

Phase Response Test

Checks phase shift between input and output signals at 1 Watt output.

Level Controls Test

Verifies gain reduction as level controls are rotated counterclockwise.

Current Limit Test

Measures the amplifier's current limit point with a 1 Ohm load.

Slew Rate & 10 kHz Square Wave Test

Measures slew rate and verifies the integrity of a 10 kHz square wave.

Crosstalk Test

Measures signal leakage between channels at 20 kHz.

Output Power Test

Measures output power at 8, 4, and 2 Ohm loads with specified THD.

Reactive Loads Test

Tests amplifier stability and safety with capacitive, inductive, and shorted loads.

ODEP Limiting Test

Verifies ODEP limiting waveform and operation in Parallel Mono Mode.

LF Protection Test

Checks if the amplifier mutes for low frequency signals.

Signal to Noise Ratio Test

Measures signal-to-noise ratio at specified conditions and bandpass filter.

Turn On Transients Test

Monitors the output for dangerous transients during amplifier turn on.

Turn Off Transients Test

Monitors the output for dangerous transients during amplifier turn off.

Intermodulation Distortion Test

Measures intermodulation distortion at different output levels using SMPTE standard.

Clipping Test

Verifies that no protective circuits activate during a clipping test.

Post Testing Procedures

Related product manuals