nVision Operation Manual Page 41
.X
Accuracy.in.Differential.Pressure.Mode:
Using the Tare function on a Dierential Pressure Measurement when two pressure modules of
the same full scale pressure range, the following accuracy statement applies:
The greater of
Full Scale Range
of Both Sensors
PSI PSI mbar inH2O % of DP Reading
30 0.0015 0.10 0.042 or 0.025%
100 0.0040 0.28 0.11 or 0.025%
300 0.015 1.0 0.42 or 0.025%
1 000 0.04 2.8 1.1 or 0.05%
3 000 0.15 10.3 4.2 or 0.05%
10 000 0.4 27.6 11.1 or 0.1%
.X
Logging.Interval
Fastest Logging Interval 10 readings per second
.X
Media.Compatibility
Liquids and gases compatible with sensor and CPF tting system:
Wrench-tight: 316 Stainless Steel
Finger-tight: 316 Stainless Steel and Viton (internal MPF o-ring)
.X
Connection
Crystal CPF System: Medium Pressure Female (MPF) (1/4” medium pressure tube system with
7/16-20 threads). See Crystal Engineering’s CPF Brochure for further information.
CPF o-ring size and material: AS568A-012, Viton 90 durometer.
CAUTION: To achieve CPF maximum allowable working pressures no o-ring substitutions
are allowed. See Crystal Engineering’s CPF brochure and CES-003 CPF Safety Guide avail-
able from the website at http://www.crystalengineering.net for further detail.