Network Terminal Installation, Operation, and Maintenance Instructions
P/N 8040375G001 4-9
C
HELIX TECHNOLOGY CORPORATION
-
TI CR
Y
OG
E
N
IC
S
Setting the RS-232 Port Baud Rates
1. Press SERVICE 9 9 to access to the local Service function of
the terminal. Press Next, then Next again to display the password
prompt.
2. Enter a password; then press Enter and Next.
3. Continue pressing Next until 9600 HOST BAUD is displayed.
4. To change the baud rate press 0 for 2400, 1 for 9600, 2 for 19200 or
3 for 38400 baud.
5. Press Next to display 9600 SERVICE BAUD.
6. Press Next to display 9600 AUX BAUD.
7. Press Next or any function key to exit the local Service function.