Why is there little or no flow in my Cytiva Laboratory Equipment?
- SSarah ReedAug 5, 2025
If you experience little or no flow in your Cytiva Laboratory Equipment, verify that the connected inlet is the one being used. Ensure liquid is supplied to the system and check the inlet containers. Confirm that the pressure control valve is open. Inspect the tubing from the inlet container for any issues like excessive length, small internal diameter, narrow sections, or blockages. Also, check the peristaltic tubing for wear, flattening, or improper tension within the pump head; re-tension or replace the tubing as needed.