DIGITAL INDICATOR
DN500N
13
5-3. Error Display and Troubleshooting
No. MESSAGE CAUSE SOLUTION
1 Err 01
It is displayed when Max. display
division/value of one division is
over 20,000
- Please re-input the value of Max display
division and one division so that Max. display
division/ value of one division is less than
10,000
4 Err 04
It is displayed when standard weight
setting is over Maximum display
division.
- Please re-input a standard weight setting
value less than Maximum display division by
using number Key.
5 Err 05
It is displayed when standard weight
setting is less than 10% of Maximum
display division.
- Please re-input standard weight setting
value less over 10% of one division.
6 Err 06
It is displayed when Amp and Gain is
too big.
- Please check if standard weight exceeds the
value you already set. If so, please set the
standard weight according to the value
correctly.
7 Err 07
It is displayed when Amp and Gain is
too small.
- Please check if standard weight is less than
the value you already set. If so, please set
the standard weight according to the value
correctly.
8 Err A
It is displayed when weight value is
fluctuated during calibration.
-Please check surroundings and remove any
vibration.
-Check defect of scale.
-Check scale connection or wiring condition.
9 Err _8
It is displayed if wrong data is
entered for F-Function
- Check the data and re-enter.