DIGITAL INDICATOR
DN500N
16
6-2. F-FUNCTION LIST
F-Function DESCRIPTION DIVISION
F00 Select Set-UpCalibration Divided into CLR and Input ENTER
F01 Set the location of decimal Point 0, 0.0, 0.00, 0.000
F02 Zero Memory Mode. Normal(0), Back UP(1)
F03 MOTION BAND Range 0, 1, 2, 3
F04 ZERO TRACKING Range 0, 1, 2, 3
F05 Set AUTO ZERO Range 0099
F06 Digital Filter Range 1∼9
F07 ZERO, TARE Key Operation Mode Steady(0), Unsteady(1)
F08 Set ZERO Key Operation Range 2%(0), 5%(1), 10%(2), 20%(3), 100%(4)
F09 Set TARE Key Operation Range 10%(0), 20%(1),50%(2),100%(3)
F10 Set HOLD function Peak-hold(0), Sample hold(1),8Sec Average Hold(2)
F11 Set External Input 0,1,2,3,4
F12 Appoint Code number 0,1,2
F13 Set S/N Key Operation 0,1
F14 Time for Hold OFF 0.0 9.9 Sec
F30 Set Serial 2 Parity Bit NO(0), ODD(1), EVEN(2)
F31 Serial 2 Communication Speed 0 9 , 115200 bps 2400 bps
F32 Serial 2 Communication Mode
0 : Stream Mode, 1 : Steady Mode, 2:PRINT KEY
F33 Serial 2 Communication Method
0:One-way Transmit Mode, 1:COMMAND MODE, 2:LCD MODE
F34 Set ID NUMBER 199
F35 Transmit Data FORMAT
0 : Basic FORMAT , 1 : Basic+Time , 2 : CAS FORMAT
F36 Select BCC 0 : BCC No Use 1 : BCC Use
F40 Set Weighing Unit Print 0:kg, 1:g, 2:ton
F41
Select Data Output with Auto Printer 0 :F80 Setting , 1 : Steady Lamp
F42 Set Printing Format 0 : Consecutive Print, 1 : Individual Print
F43
Set removing printer subtotal Memory
0 : Remove Subtotal Grand 1 : Auto remove
F44 Set paper feeding 1Count increase= 1 Line increase
F45 Set print line interval
Control the interval 1 line to the next line
F46 Subtotal Print Mode
0 : Print weight value, 1 : Print Max. Min. Average
F47 Select PRINT Font 0 : Korean 1 : English
F48 PRINT Delay Time 09.9
F52 Select Key Tare Operation 0: Key Tare No Use,1: Key Tare Useable
F77 FUNCTION AUTO SETTING AUTO FUNCTI0N SETTING
F80 Set NEAR ZERO(EMPTY) Range x x x x x x
F81 Set Zero Display Range x x x x x x
F89
Confirm Calibration SPAN Constant Value
x. x x x x x
F90
Confirm & Change Date (Year,Month,Date) x x. x x. x x
F91 Confirm & Change Time (Hour,Min,Sec) x x. x x. x x