12  DKRCI.PD.SC0.C3.02 / 520H5687  © Danfoss A/S (AC-MCI/MWA), 2012-05
Liquid Level Sensor, AKS 4100/4100U
•   Connect the device to the power supply 
  (see the section "Electrical installation/ 
 connection".
•   Press   3 times.
•   Press   
•   Press 
 
or 
 
to select either SINGLE 
  or COAXIAL. Press   to conrm. 
•   Press 
 
to change the PROBE LENGTH. 
 Press 
 
to change the position of the  
 cursor.  
 Press 
 
to decrease the value or 
 
to 
  increase the value. 
 Press   to conrm. 
AKS 4100
QUICK SETUP ?
YES               NO
AKS 4100
PROBE TYPE
SINGLE CABLE
AKS 4100
PROBE  LENGTH
05000 mm
AKS 4100
SCALE 4 mA
04946 mm
AKS 4100
SCALE 20 mA
00070 mm
•   Press 
 
to change of SCALE 4 mA. 
 Press 
 
to change the cursor position. 
 Press 
 
to decrease the value or 
 
to 
  increase the value. 
 Press   to conrm. 
•   Press 
 
to change of SCALE 20 mA. 
 Press 
 
to change the cursor position. 
 Press 
 
to decrease the value or 
 
to 
  increase the value. 
 Press   to conrm.
•   Wait for QUICK SETUP to complete    
  8-second timeout
AKS 4100
QUICK SETUP 
COMPLETED IN 8
AKS 4100
1.0.0
QUICK START
AKS 4100
1.0.0
STORE NO
AKS 4100
DISTANCE
5000 mm
•   Press   to conrm.
•   Press 
 
or 
 
to select either  
  STORE NO or STORE YES.  
 Press   to conrm. 
  Default screen appears:
  Quick Setup completed
  You have the possibility of checking 
  your settings by pressing 
 
twice.
 Press   
 
  to return to 
  default screen.
Quick Setup (all values below are only examples)
QUICK SETUP ?
AKS 4100
QUICK SETUP ?
YES                     NO
Note: 
The signal converter can be programmed with or without mechanical process connector assembled.