ISM-3 EN User Manual 15
Use to display the next sub menu.
The setup menu SET.
1. Press when SET. appears on the display.
2. Use and to set required value.
3. Press and hold for 3 seconds to change the value. (for typing in values, see "Set-up of alarm
limits" on page 13).
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
FLO.L 100 Flow alarm limit. Alarm limit for min. flow in ml/min.
FLO.H 150 Flow alarm limit. Alarm limit for max. flow in ml/min.
DEL. 3 Gas flushing time in seconds before activation of alarms.
Valid area 0 to 250.
PA. 1001 Setting up of options, relay function and external measurement control.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
When typing in values for FLO.L and FLO.H, error ER.1 appears, if FLO.L is set to a higher value than
FLO.H.
Diagnostic menu DIA.
1. Press when DIA. appears in the display.
2. Use and to set required value. For a permanent readout of the chosen value press and hold
for about 3 seconds. The value wil now be displayed until you press , or again.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
EMF. 165.0 Present sensor EMF in mV.
TPT.1 33 Present internal ISM-3 temperature in °C.
TPT.2 0 Present difference in heater temperature in °C. Difference between present
temperature and ideal temperature.
OFFS -0.12 Sensor offset. Result of latest off set adjustment in mV.
TI.PO 1234 Hour counter in number of 10 hours. Number of hours in which ISM-3 has
been switched on.
TI.CA 678 Hour counter in numbers of 10 hours. Number of hours since the last
calibration of ISM-3.
TI.FL 983 Hour counter in number of 10 hours. Number of hours in which the flow has
passed through the sensor (flow >30ml/min).
PRG. 1.0 Installed program version.
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Calibration menu CAL.
See the chapter about calibration