16 User Manual EN ISM-3
Setting up of options, relay and external measurement control
In the set-up menu the PA. parameter is defined as follows:
The first digit (from the left) determines from where and how the measurement can be started and
stopped.
The second digit must be 0 and is reserved for future use.
The third digit determines how the relay works.
The last digit indicates if there has been installed options as current output/ voltage output.
0 : Press Measurement activated only via button (manual control)
1 : Press/Ext. Measurement controlled via button and external
measuring control signal
2 : Ext. Measurement activated only via external measuring
control signal.
3 : None Measurment is constantly connected and cannot be
disconnected
0 : No current output / voltage output
1 : Current output (0-20mA/4-20mA) or
voltage output (0-10Volt/2-10volt)
Measurement
0
Current
output
Relay
0
Off
Off
Off
Off
Off
Off
ISM-3 mode
When heating
System error
High oxygen alarm
Low oxygen alarm
Flowalarm
Measurement off
1
-
Off
Off
Off
Off
-
2
On
On
On
On
On
On
3
-
On
On
On
On
-
Do not change
Current output / voltage output
Effect of ISM-3 relay
Measurement controlled by: