Dansensor A/S COPYRIGHT © PBI-200114-D 04/2013
ISM-3 EN User Manual 21
7. Options
Current output/voltage output
ISM-3 can be equipped with an current output/voltage output. From the the set-up menu the output can
be set to the ranges 0-20mA or 4-20mA/0-10V or 2-10V respectively.
The current output is enabled from the
PA. parameter in the set-up menu. When PA. parameter digit 4 is
set to “1”, 3 menu items are automatically added to the set-up menu.
CU.0 - - - - O
2
value which must correspond to 0mA (0 V)
CU.4 0% O
2
value which must correspond to 4mA (2 V)
CU.20 10% O
2
value which must correspond to 20mA/10V
Only one of the parameters
CU.0 and CU.4 can be se, the other one is displayed as “- - - -”.
If identical values are entered for the 0 and 20mA or the 4 and 20 mA values, error code
er. 1 will be
displayed.
DEL
Pa.
cu. 0
cu. 4
Cu.20