PBI-200114-D 04/2013 COPYRIGHT © Dansensor A/S
24 User Manual EN ISM-3
The display shows : er. 4
An attempt has been
made to change
calibration gas No. 1
(20.946 %O
2
). This gas is
reserved for atmospheric
air.
Select one of the other calibration gas entries
(where readout is 0 ppm) for entry of a new
calibration gas.
The display shows:
er. 5
An attempt has been
made to enter two
calibration gases with the
same value
Find the calibration gas with the same value as
the one you tried to enter and calibrate
The display shows:
er. 6
An attempt has been
made to calibrate with a 0
ppm gas
Type in a gas that is not 0.
The dispaly shows:
er. 7
The result of the offset
adjustment is too high
(> 10mV)
Check that ISM-3 is only supplied with
atmospheric air during offset adjustment
The display shows:
er. 8
An attempt has been
made to calibrate with an
other gas than 20.946%
O
2
(atmospheric air) after
offset adjustment has
been carried out
When offset adjustment has been carried out,
all calibrations are locked. See "Calibration" on
page 19.
The display shows :
er.9
An attempt has been
made to calibrate or
adjust the offset where
measuremet has been
disabled or where warm-
up period has not been
completed.
See "Calibration" on page 19 and/or "Offset
adjustment" on page 20.
The display shows:
er.30
Hardware error. The mains
supply line failure signal is
constantly high
Switch off ISM-3, wait for 10 seconds and
switch it on again. If an error is still indicated,
ISM-3 should be returned for repair.
It is possible to continue by pressing .
However, the mains supply unit is defective and
should be replaced / repaired (possibly a bad
wire connection).
The display shows:
er.31
The last time ISM-3 was
switched off, the line
failure signal from the
mains supply unit was not
detected. If this error
occurs in connection with
each start, there is a
hardware error
Switch off ISM-3, wait for ten seconds and
switch it on again. If an error is still indicated,
ISM-3 should be returned for repair.
It is possible to continue by pressing .
However, the mains supply unit is defective and
should be replaced / repaired (possibly a bad
wire connection).
Problem Possible cause Possible solution