IMPORTANT! Do not provide these setup instructions to the patient. Only qualified
medical professionals should access these device settings.
Follow these instructions to access the device settings and prepare the device for
patient use. All settings have pre-set values as shown below.
Other Settings
Filter Cleaning Reminder – Displays a reminder to the patient to clean the filter every 50 hours of use, or approximately every 10 days. This
feature is controlled by the notifications setting and is preset to enabled.
Other setup information
This device features automatic altitude adjustment for altitudes up to 2743m (9000ft) and needs no manual adjustment.
This device features a pressure sensor therefore no ne pressure adjustments are needed.
The SmartLink Data Management System may be used for additional reminders and notications.
Clear compliance meter by pressing Down key for eight seconds with compliance meter shown (value will go to zero).
All device settings are immediately updated in the device as you make adjustments. When nished press the ON/OFF button once to exit the
Clinical Setup Menu.
Use the LEFT button
and RIGHT button to
navigate through the
menu.
Use the UP button and
DOWN button to adjust
the settings.
First - With the device powered
and turned off, press and hold
the DOWN and DELAY buttons.
Menu Item Description
Factory
Default
Clinical Menu
Language
Selection
Select device language. English
Set
Pressure
Set to CPAP pressure prescribed by doctor.
Range: 3-20 cmH
2
O
10.0 cmH
2
0
Delay
Time
Select period of time to stay at delay pressure. This delay is patient activated.
Range: 0-45 minutes, increments of 5
20 min
Delay
Pressure
If prescribed, set to a level that is comfortable for the patient to fall asleep. This level of pressure is maintained
for the first portion of the delay time and rises to the set pressure in the last 10 minutes of the delay time.
Range: 3 cmH
2
O to CPAP, increments of 0.5
4.0 cmH
2
0
Pressure
Unit
The device operating pressure may be displayed in one of three pressure units: cmH
2
O, hPa or mbar. cmH
2
0
Compliance
Meter
Cumulative hours of “while breathing” time.
N/A
Enable
Menu
Press the button to enter. Reference the Enable Submenu section for details.
N/A
Lockout
Menu
Press the button to enter. Reference the Lockout Submenu section for details.
N/A
Enable Submenu
Auto-ON
This option controls the Auto-ON feature, which automatically starts the ow of air after breathing once or
twice into the mask
Enabled
Auto-OFF
This option controls the Auto-OFF feature, which automatically stops the ow of air after taking off the mask.
Disabled
Notifications
Setting to enable mask/filter check notification.
Enabled
Exit
Submenu
Press the button to exit and return to Clinical Menu.
N/A
Lockout Submenu
Delay Time
Lockout
Used to allow patient access to CPAP delay time setting.
Range: Locked, Unlocked
Unlocked
Exit
Submenu
Press the button to exit and return to Clinical Menu.
N/A
Quick Set-Up Guide
IntelliPAP™ Standard SleepCube™ Standard
EN
DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881
Sunrise Medical Ltd. • Sunrise Business Park • High Street • Wollaston, West Midlands • DY8 4PS England • 44-138-444-6688
©2007 DeVilbiss Healthcare 12.07
DV51D-125 Rev. A
Second - While holding the
two buttons, press the ON/OFF
button. The display will show
“Clinical Menu”. You can now
access the Clinical Setup Menu.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the DeVilbiss IntelliPAP Standard and is the answer not in the manual?

DeVilbiss IntelliPAP Standard Specifications

General IconGeneral
BrandDeVilbiss
ModelIntelliPAP Standard
CategoryMedical Equipment
LanguageEnglish

Related product manuals