EN - 16 SE-DV57-1
Dynamic Short-Term Pressure Accuracy with and without DV5HH Humidifier (Per
ISO 17510-1:2007)
Waveform Volume [mL] Breath Rate [min-1]
Set Pressure
[cmH
2
O]
Pressure Accuracy
[Pk-Pk cmH
2
O]
Sinusoidal
cyclewithI:E=
1:1
500 10, 15, 20
3.0 0.5
9.0 0.5
14.0 1.0
20.0 1.0
25.0 1.0
Maximum Flow-Rate (per ISO 17510-1:2007)
Test Pressures
3.0 cmH
2
O 9.0 cmH
2
O 14.0 cmH
2
O 20.0 cmH
2
O 25.0 cmH
2
O
Measured pressure at the patient
connection port (cmH
2
O)
3.0 8.9 13.0 19.0 24.0
Averageowatthepatientconnectionport
(l/min)
95.0 160.5 172.2 158.5 139.7
Long Term Static Pressure Accuracy ................................................................................................................. +/-0.5cmH
2
O
Additional Specifications
Equipment classification with respect to protection from electric shock ........................................................................ Class II
Degree of protection from electric shock ................................................................................................. Type BF Applied Part
Degree of protection against ingress of liquids ...................................................................................... IPXI Drip-proof vertical
Mode of operation .................................................................................................................................................... Continuous
Equipmentnotsuitableforuseinthepresenceofaammableanestheticmixturewithair,oxygenornitrousoxide.