EN - 15SE-DV57-1
Issue Possible Cause Remedy
Air from ow
generator seems to
be too warm.
1.Airltersaredirty. 1.Cleanlters.
2. Air inlet port is blocked. 2. Unblock air inlet.
3. Room temperature is too high. 3. Lower room temperature.
4. Device is located near a heat source. 4. Locate device away from heat source.
5.Humidier(optional)heatersettingistoo
high.
5. Lower heater setting.
Nasal, sinus or ear
pain, runny nose
Youmayhaveareactiontoairowpressure. Discontinue use and contact physician.
Heater setting is
not visible on
display
1.Theowgeneratorisnotfullyseatedonto
thehumidiercradle.
1. Ensure proper contact is made (Fig. 3).
2. Your CPAP device is not equipped with the
optionalhumidier.
Bilevel changes
pressure too early
or too late (not
following breathing)
1. Trigger Sensitivity is not set correctly. 1.Adjusttheinspiratoryandexpiratorytrigger
sensitivitytoadjusttriggeringforthepatient.
2. Backup rate (time) is too high, the patient is
breathing slower than the backup rate.
2. Reduce Backup Rate setting.
SPECIFICATIONS
CPAP
Size ........................................................................................................... 4.2”H x 6.5”W x 6.9”D (10.7cm x 16.5cm x 17.5cm)
Weight ............................................................................................................................................................. 1.22 kg (2.7 lbs.)
Electrical Requirements AC ......................................................................................................................100-240V~, 50/60 Hz
Electrical Requirements DC .................................................................................................................. 10.4 - 15 VDC, 5 Amps
Maximum Power Consumption ......................................................65 watts max from AC power source (flow generator only)
Bilevel Pressure Range ........................................................................................................................................... 3-25 cmH
2
O
Operating Temperature Range .......................................................................................................41˚Fto104˚F(5˚Cto40˚C)
Operating Humidity Range .......................................................................................................0% to 95% RH non-condensing
Operating Atmospheric Conditions ..........................................................................................Sea level to 8,500 feet (2600 m)
Storage & Transportation Temperature Range ........................................................................-40˚Fto+158˚F(-40˚Cto70˚C)
Storage & Transportation Humidity Range ...............................................................................0% to 95% RH non-condensing
Maximum Limited Pressure ...........................................................................................................30 cmH
2
O under normal use
Sound Pressure Level (tested per ISO 17510) ...............................................................................................................26 dBA
Sound Power Level .........................................................................................................................................................38 dBA
SmartCode
®
Technology is protected under US patent 8649510 and patent HK 1172119
Humidifier
Humidity Output (in the operating flow range): ................................................................................................ ≥10mgH
2
O/l air
Size: ......................................................................................................2.6” H x 6.3” W x 8.4” D (6.6 cm x 16.0 cm x 21.3 cm)
Weight: ........................................................................................................................................................1.75 lbs. (0.794 Kg)
Power Consumption: .................................................................................................................................................... 85 Watts
DV5HH Electrical Requirements ................................................................................................................100-240VAC, 0.95 A
Heater Plate Thermal Cutout:..............................................................................................................................305˚F(152˚C)
Power/Temperature Control: ......................................................................................Setting of 1 (minimum) to 10 (maximum)
Heater Plate Temperature: .................................................................................84˚Fto149˚F(approx.29˚Ctoapprox.65˚C)
Operating Temperature Range: ......................................................................................................41˚Fto104˚F(5˚Cto40˚C)
Operating Humidity Range: ....................................................................................................... 0 to 95% R.H. non-condensing
Operating Atmospheric Pressure Range: ........................................................................................................70.0 – 106.0 kPA
Filter Specifications
Standard Filter ......................................................................................................................................... > 3.0 micron particles
Optional Fine Particle Filter ..................................................................................................................... > 0.3 micron particles