Operator's Manual DS7A - Constant Current Stimulator
FAULT LED - Amber, illuminated on (i) sensed over-current internally (ii) over-
current to the patient or (iii) Trigger frequency “Too-fast” in alternating
polarity mode. The unit is latched in this condition until reset by the Output
On/Off switch being placed in the OFF position.
POWER ON LED - Green, illuminated for power on
External Fuses
Note these should ONLY be changed by qualified persons
100 - 120 V mains T 250 mA L 5 x 20 mm, 250 V
200 - 240 V mains T 125 mA L 5 x 20 mm, 250 V
Internal Fuses
Note these should NOT be changed by non-qualified persons
Fuse Reference F3 and F4 – TE5 400 mA 125 V Antisurge.
Dimensions
Size: 255 x 100 x 225 mm (d x h x w)
270 x 110 x 225 mm (d x h x w) - over controls and feet
Weight: 2.1 kg (approx.)
Temperature
Operating Range 10°C to 40°C 30 to 75%, non-condensing
Storage Range -40°C to 70°C 10 to 100%, non-condensing
Transport Range -40°C to 70°C 10 to 100%, non-condensing
Other
Power: 100 - 120V or 200 - 240V @ 47-63 Hz
Rating: <30 VA
Case Material UL94 V-0 Flame Retardant