9
If the rotor needs cleaning, clean with cloth or sponge moistened with a neutral detergent solution.
Sterilize the centrifuge by wiping with a cloth moistened with 70% ethanol solution.
3Drive shaft
We recommend regular maintenance for drive shaft. You can wipe the drive shaft with soft cloth, and
then apply a thin coat of silicon grease.
4Door
Clean and sterilize the door using the same method as the section1above.
5Rotor
For sample leaks in the rotor, rinse the rotor with waterand dry it completely.
The rotor should be checked every 3 months to ensure the tube and rotor holes keep are clean.
8.2 Rotor Installation
1Installation
2Adjustment
CAUTION
Do not directly pour water, neutral detergent or disinfectant solution into the rotor chamber, otherwise
fluids may leak into the drive units and cause corrosion or deterioration to the bearings.
Observe here!